CALIBRATION LABORATORY

for flow meters DN10 to DN200

  • paddle type / rotating vane
  • electro-magnetic
  • ultrasonic

Technical specification of calibration facility:

type
MQ-T-S 40150
for flow meters of DN
10 to 200mm
flow rate range
0,150 to 150m3/h
fixation device lenght
210mm
fixation
pneumatic, mechanically safeguarded
calibration reservoir capacity
5m3
waterpump
WILO, 23kW
el.-mag. noise resistance EMC compatibility
according to OIML D11
method of measurment
volumentric comparison
etalons
YOKOGAWA AXF 015G DN15
YOKOGAWA AXF 015G DN25
YOKOGAWA AXF 015G DN80
combined uncertainty
Uc = 0,17%
controlling and evaluating system
ENBRA
control method
using PC: automatic
semiautomatic
manual
other measuring instruments
3x thermomethers Pt100
2x pressure gauge